Compact gas chromatograph in Pakistan: LDetek MultiDetek-Lite MD-Lite 4U rackmount GC for trace impurities in industrial, specialty and electronic gases
LDetek MD-Lite: compact 4U rackmount GC, up to 3 ovens and 2 detectors (PED / TCD / FID), ppb to %, up to 8 × 4–20 mA (datasheet).

Industrial gas fillers, air separation plants, semiconductor and electronics fabs and food-grade CO2 and nitrogen users all face the same question: how many impurities are actually in the gas, and at what level? A single-gas analyser answers one part of that question. When you need several trace impurities, a purity figure and sometimes trace oxygen and moisture as well, a gas chromatograph built around the job is usually the cleaner answer. The challenge is that a full process GC can be more instrument than a straightforward duty needs.

This note looks at the LDetek MultiDetek-Lite (MD-Lite) compact gas chromatograph, LDetek's most compact GC series in a 4U rackmount. Figures come from the MD-Lite datasheet (MULTIDETEK-LITE_V1_EN_DATASHEET_0524) hosted on our product page and from the current Tetracon product pages. Where we give general engineering advice it is labelled as general. Nothing here goes beyond the manufacturer's published data. Tetracon (Pvt.) Ltd. is the sole authorized agent for LDetek in Pakistan.

Why a compact GC for trace impurities?

  • Several impurities, one instrument. The MD-Lite takes up to three isothermal ovens for GC columns and up to two chromatography detectors in one unit, with a 4–20 mA output per impurity (up to 8 outputs) (datasheet).
  • ppb to percent. LDetek describes a large measurement scale from ppb to %, combining a plasma emission detector (PED) for trace levels with a thermal conductivity detector (TCD) for percent-level purity (datasheet).
  • Same platform, smaller box. The MD-Lite is based on the MultiDetek3 platform and LDetek states it offers the same performance in a smaller environment, as an alternative when the application needs less complexity (datasheet).
  • Cost-saving. The datasheet positions the 4U configuration as a cost-saving solution for multiple trace-impurity applications (datasheet).
  • Online O2 and H2O in the same unit. Optional online detectors add trace oxygen (Ntron SenzTx, zirconia or electrochemical) or trace moisture (Michell quartz crystal QMA) alongside the GC channels (datasheet).

How the MD-Lite measures

Sample gas is separated on GC columns in isothermal ovens and the separated components are measured by the chromatography detectors. Electronic flow controllers manage the sample and carrier flows, and an internal sampling system switches between span and sample (datasheet).

  • PlasmaDetek (PED). A plasma emission detector with selective optics and interference filters, matched to the chromatography columns for the application. It works with helium, argon or nitrogen carrier (application dependent) and is aimed at trace ppb/ppm impurities (datasheet).
  • TCD. Measures sample gas purity at percent level, or several impurities at high ppm levels with suitable columns. Combined with the PED in one unit, LDetek quotes coverage from 1 ppb (PED) to 100 % (TCD) (datasheet).
  • FID. Available when carbons detection by flame ionization is required. LDetek notes that a PED with nitrogen carrier is widely used in its systems to replace the FID without a flammable air/H2 mixture (datasheet).
  • SenzTx (Ntron) and QMA (Michell). Optional online sensors for trace O2 and trace H2O. The QMA module self-calibrates against an internal permeation-tube moisture generator and a purified dry-gas reference (datasheet).

LDetek MD-Lite at a glance

ItemFrom the datasheet
FormatCompact gas chromatograph, 4U rackmount
Ovens / detectorsUp to 3 isothermal ovens; up to 2 chromatography detectors (PED / TCD / FID)
PED (PlasmaDetek)Permanent and noble gases, sulfurs, ammonia, hydrocarbons, aldehydes, BTEX, alcohols…; up to 5000 ppm available; LDL 100 ppt (application dependent); sensor life 10 years
TCDUp to 100 % available; LDL 1 ppm (application dependent); life 3–5 years
FIDHydrocarbons, CO, CO2; up to 100 % available; LDL 1 ppb (application dependent); life 3–5 years
SenzTx O2 (Ntron)Zirconia up to 96 %, LDL 1 ppm; electrochemical up to 25 %, LDL 0.5 ppm
QMA H2O (Michell)0–10 ppm (resolution 100 ppb), 0–100 ppm or 0–1000 ppm (resolution 1 ppm), up to 2000 ppm available; LDL 20 ppb
Sample gasAny UHP gases, gas mixtures and ambient air
Operating temperature / sample temperature5–45 °C / 0–100 °C
Sample pressure / outlet5–30 psig (0.35–2 barg) / atmospheric
Sample flow100 ml/min (QMA 300–500 ml/min)
Carrier gasHe, Ar or N2 (PED); He, Ar, N2 or H2 (TCD, FID), application dependent; 80–100 psig (5.5–7 barg)
Fittings1/8" or 1/4" compression or face seal
Standard I/OMCU system, 5" touchscreen, 4 × 4–20 mA, 4 × dry contacts, remote control via Ethernet
OptionsModbus via RS485, RS232 or Ethernet; Profibus; 4 extra 4–20 mA outputs; 4 extra dry contacts
Supply / power110 V AC or 220 V AC, 50–60 Hz; maximum 250 W depending on configuration
Weight20–30 kg (45–65 lb) depending on configuration
CertificationEMC: IEC 61000-4-x immunity; CISPR 32 and FCC Part 15 Subpart B emissions
LDetek MultiDetek-Lite MD-Lite compact 4U rackmount gas chromatograph with key specifications from the datasheet
LDetek MD-Lite: up to 3 ovens and 2 GC detectors in a 4U rack, ppb to %, 4–20 mA per impurity (datasheet).
LDetek MD-Lite detector options compared: PED, TCD, FID, SenzTx and QMA ranges and detection limits
MD-Lite detector options: PED, TCD, FID, SenzTx O2 and QMA H2O ranges and detection limits (datasheet).

Choosing the detector combination

  • Trace impurities in a pure gas (for example ppb or ppm N2, O2, CO, CO2 or hydrocarbons in argon, helium or nitrogen): the PED is the detector built for this, with application-dependent detection limits down to 100 ppt (datasheet; general).
  • Percent-level purity or mixture composition: add the TCD, which covers up to 100 %; LDetek pairs PED and TCD for full coverage from ppb to percent (datasheet).
  • Hydrocarbons where an FID is specified: the FID remains available, but consider whether a PED on nitrogen carrier can do the job without bringing an H2/air flame into the lab or analyser house (datasheet; general).
  • Trace oxygen online: choose the SenzTx zirconia (up to 96 %, LDL 1 ppm) or electrochemical (up to 25 %, LDL 0.5 ppm) module; note the electrochemical sensor life is given as 1 year against 3–5 years for zirconia (datasheet).
  • Trace moisture: the QMA module covers 0–10 ppm up to 2000 ppm with an LDL of 20 ppb, and needs 300–500 ml/min of sample rather than 100 ml/min (datasheet).

Practical notes for a reliable installation

  • Plan the carrier gas. GC detectors need a carrier supply at 80–100 psig (5.5–7 barg), and the carrier choice is application dependent; agree it with the application study before ordering cylinders or a generator (datasheet; general).
  • Control sample pressure. The sample inlet expects 5–30 psig (0.35–2 barg) with atmospheric outlet; high-pressure cylinder or pipeline samples need a regulator, and vents should go somewhere safe (datasheet; general).
  • Keep the sample path clean. For ppb work use clean stainless tubing with 1/8" or 1/4" compression or face-seal fittings, minimise dead volume and purge new lines before trusting a result (datasheet fittings; general practice).
  • Mind the room. The operating temperature is 5–45 °C, so a 4U rack in an air-conditioned analyser room or lab suits most Pakistani sites better than an unconditioned shelter (datasheet; general).
  • Size the power. Allow up to 250 W depending on the sensors and options fitted, on 110 or 220 V AC (datasheet).
  • Hazardous areas are a different product. The MD-Lite is a 4U rackmount for safe areas; for Zone 1 or Zone 2 installation look at the purged MultiDetek3 EX instead (product pages; general).

Where it fits in Pakistan

  • Industrial gas producers and fillers certifying argon, nitrogen, oxygen, helium and mixtures before dispatch (datasheet application; general).
  • Energy users checking gas quality where a full process GC would be oversized (datasheet application; general).
  • Semiconductor and electronics assembly lines that specify ppb impurities in bulk and specialty gases (datasheet application; general).
  • Food and beverage plants checking CO2 and nitrogen used for carbonation and packaging (datasheet application; general).
  • Environmental and laboratory users who want a compact, rack-mounted GC with remote control over Ethernet (datasheet; general).

Related products and when to choose them

DutyProductKey figure (datasheet / product page)
Several trace impurities plus purity in a compact safe-area rackLDetek MD-Lite4U rackmount; up to 3 ovens, 2 GC detectors; ppb to %; up to 8 × 4–20 mA (datasheet)
GC in a Zone 1 / Zone 2 hazardous areaLDetek MultiDetek3 EXATEX / IECEx purged stainless enclosure; up to 6 ovens, up to 10 valves (product page)
Only trace N2 in argon or helium, without a GCLDetek LD8001+PED, ppb-level trace nitrogen, 3U cabinet, 4–20 mA (product page)
Trace N2 and O2 for gloveboxes and AMLDetek LDSENZPED for nitrogen, SenzTx ZR or EC for oxygen (product page)
Integrating analysers, stream selector and purifier in one cabinetLDetek LDRack19" cabinets 9U to 42U, LDGSS stream selector 1–20 streams (product page)
Standalone trace moisture in process gasMichell QMA601Quartz crystal, 0.02–100 ppmV trending to 2000 ppmV (product page)
Standalone trace or percent oxygen transmitterNtron SenzTxZR 0–1000 ppmV to 0–100 %; EC 0–1000 ppmV to 0–25 % (product page)

For the hazardous-area version of the same platform, read our MultiDetek3 EX note; for single-impurity alternatives see the trace nitrogen note and the LDSENZ note, and for moisture the quartz-crystal moisture note. The wider range is on the Gas Analysis and Oxygen Analysis category pages, our selection guides, and the LDetek, Ntron and Michell Instruments manufacturer pages.

What to send with an enquiry

  • The background gas (argon, nitrogen, helium, oxygen, hydrogen, CO2, a mixture or ambient air) and its expected purity.
  • Every impurity you need to measure, with the expected level and the specification limit (ppb, ppm or %).
  • Whether trace O2 and/or trace H2O must be measured online in the same unit.
  • Sample pressure and temperature at the take-off point, number of sample streams, and whether a span gas is available.
  • Carrier gases available on site, the supply voltage (110 or 220 V AC) and where the 4U rack will be installed.
  • Outputs needed: number of 4–20 mA signals and dry contacts, Modbus (RS485, RS232 or Ethernet) or Profibus.

FAQ

What is the LDetek MD-Lite?

The MultiDetek-Lite (MD-Lite) is LDetek's most compact gas chromatograph series, in a 4U rackmount. It is based on the MultiDetek3 platform and combines up to three isothermal ovens and up to two chromatography detectors for multiple trace-impurity applications.

Which detectors can it use?

Chromatography detectors PED (PlasmaDetek), TCD and FID, up to two in one unit, plus optional online detectors: Ntron SenzTx zirconia or electrochemical sensors for trace O2 and a Michell quartz crystal (QMA) sensor for trace H2O.

How low can it measure?

Detection limits are application dependent. The datasheet gives 100 ppt for the PED, 1 ppb for the FID and 1 ppm for the TCD, 1 ppm (zirconia) or 0.5 ppm (electrochemical) for oxygen and 20 ppb for the QMA moisture sensor.

What outputs and communications does it have?

Standard: 5-inch touchscreen, 4 × 4–20 mA outputs, 4 × dry contacts and remote control via Ethernet. Options add Modbus via RS485, RS232 or Ethernet, Profibus, and 4 extra 4–20 mA outputs and 4 extra dry contacts.

Can the MD-Lite be installed in a hazardous area?

The MD-Lite is a 4U rackmount unit with EMC certification. For Zone 1 or Zone 2 locations, LDetek offers the MultiDetek3 EX in an ATEX/IECEx purged enclosure.

Need to measure several trace impurities in an industrial, specialty or electronic gas in Pakistan? Contact Tetracon with the background gas, impurities and limits, sample conditions and the outputs you need. You can also browse the LDetek MD-Lite product page and the LDetek manufacturer page.