Home About Us Manufacturers Products Services Clients Downloads Blog Guides FAQ Contact
MI Michell Instruments Ltd.

QMA601 Low-Range Process Moisture Analyzer

Fast, high-precision quartz crystal microbalance trace moisture analyzer for hazardous areas.

ModelQMA601
SKUMI-QMA601
CategoryDew Point & Moisture

Key Features

  • Low-range measurement 0.02–100 ppmV (trending to 2000 ppmV)
  • Accuracy ±0.03 ppmV or ±10 % of reading, whichever is greater
  • Built-in verification via internal moisture generator
  • Maintenance free for 3 years
  • Suitable for natural gas with up to 20 % hydrogen
  • IECEx, ATEX, UKCA, TR CU Ex (Ex d); cQPSus explosion proof
  • 7" color touch-screen HMI — no hot-work permit required
  • 2 × 4–20 mA or 1–5 V outputs, Modbus RTU (RS485) and Modbus TCP
  • 21 predefined and 3 user-defined carrier gases; 14 internal alarms

Overview

The QMA601 is a next-generation advanced quartz crystal microbalance analyzer designed for reliable, fast and accurate measurement of trace moisture in process applications where moisture must be kept as low as possible. It measures from 0.02 to 100 ppmV with built-in verification against an internal moisture generator, and is suitable for natural gas containing up to 20 % hydrogen with no further modification.

Working Principle

A quartz crystal sensitised with a hygroscopic film changes oscillation frequency as water molecules are adsorbed. The analyzer switches between sample and reference gas and measures the frequency change; automatic calibration uses an internal moisture generator (nominally 0.5 ppmV) traceable to NPL and NIST.

Benefits

  • Fast response to sample moisture changes
  • Low maintenance (3 years maintenance free)
  • Easy integration into SCADA via Modbus RTU/TCP

Technical Specifications

Measurement
TechnologyFast-response quartz crystal microbalance
Range0.02–100 ppmV, trending to 2000 ppmV
Accuracy±0.03 ppmV or ±10 % of reading, whichever is greater
Repeatability±5 % of reading (0.3–100 ppmV)
Limit of Detection0.02 ppmV
Sensitivity0.01 ppmV or 1 % of reading
Operating Conditions
Inlet Pressure2 barg
Outlet Pressure1 barg
Sample Flow300 ml/min total
Sample Gas Temperature0…+100 °C
Electrical
Supply85–264 V AC 47/63 Hz or 24 V DC (24 V DC only for cQPSus)
Analog2 × 4–20 mA or 1–5 V (selectable)
DigitalModbus RTU (RS485), Modbus TCP
Alarms1 system + 3 process alarms, volt-free changeover
Mechanical
EnclosureGUB flameproof Ex d, cast copper-free aluminum, IP66 / NEMA 4
Weight35 kg without sampling system
Hazardous AreaATEX/UKCA/IECEx Ex db IIB+H2 T6 Gb; cQPSus Class I Div 1 Groups B, C, D T6

Downloads

Datasheet (PDF)PDF

Frequently Asked Questions

What warranty is offered on this product?

All Tetracon-supplied instruments carry the original manufacturer warranty (typically 12–24 months) plus first-line local support from our Islamabad service centre.

Do you provide installation and commissioning?

Yes. Our manufacturer-trained engineers offer on-site installation, commissioning and startup training anywhere in Pakistan.

Are spares and calibration available locally?

Common spares are held in local stock. Calibration is available at our lab or on-site with NIST-traceable reference standards.

How long is the typical lead time?

Stocked items ship within 3–5 working days; engineered configurations typically 6–10 weeks depending on manufacturer schedule.

Request a Quote — QMA601 Low-Range Process Moisture Analyzer

Complete the form below and one of our engineers will respond within 1 working day with pricing and lead time. Prefer email? Contact our sales team.