Condumax II Hydrocarbon Dew-Point Analyzer
Fully automatic, chilled-mirror hydrocarbon and water dew-point analyzer for natural gas quality.

Fast, high-precision quartz crystal microbalance trace moisture analyzer for hazardous areas.
The QMA601 is a next-generation advanced quartz crystal microbalance analyzer designed for reliable, fast and accurate measurement of trace moisture in process applications where moisture must be kept as low as possible. It measures from 0.02 to 100 ppmV with built-in verification against an internal moisture generator, and is suitable for natural gas containing up to 20 % hydrogen with no further modification.
A quartz crystal sensitised with a hygroscopic film changes oscillation frequency as water molecules are adsorbed. The analyzer switches between sample and reference gas and measures the frequency change; automatic calibration uses an internal moisture generator (nominally 0.5 ppmV) traceable to NPL and NIST.
| Technology | Fast-response quartz crystal microbalance |
|---|---|
| Range | 0.02–100 ppmV, trending to 2000 ppmV |
| Accuracy | ±0.03 ppmV or ±10 % of reading, whichever is greater |
| Repeatability | ±5 % of reading (0.3–100 ppmV) |
| Limit of Detection | 0.02 ppmV |
| Sensitivity | 0.01 ppmV or 1 % of reading |
| Inlet Pressure | 2 barg |
|---|---|
| Outlet Pressure | 1 barg |
| Sample Flow | 300 ml/min total |
| Sample Gas Temperature | 0…+100 °C |
| Supply | 85–264 V AC 47/63 Hz or 24 V DC (24 V DC only for cQPSus) |
|---|---|
| Analog | 2 × 4–20 mA or 1–5 V (selectable) |
| Digital | Modbus RTU (RS485), Modbus TCP |
| Alarms | 1 system + 3 process alarms, volt-free changeover |
| Enclosure | GUB flameproof Ex d, cast copper-free aluminum, IP66 / NEMA 4 |
|---|---|
| Weight | 35 kg without sampling system |
| Hazardous Area | ATEX/UKCA/IECEx Ex db IIB+H2 T6 Gb; cQPSus Class I Div 1 Groups B, C, D T6 |
All Tetracon-supplied instruments carry the original manufacturer warranty (typically 12–24 months) plus first-line local support from our Islamabad service centre.
Yes. Our manufacturer-trained engineers offer on-site installation, commissioning and startup training anywhere in Pakistan.
Common spares are held in local stock. Calibration is available at our lab or on-site with NIST-traceable reference standards.
Stocked items ship within 3–5 working days; engineered configurations typically 6–10 weeks depending on manufacturer schedule.
Complete the form below and one of our engineers will respond within 1 working day with pricing and lead time. Prefer email? Contact our sales team.