Relative humidity is one of the quieter process variables in Pakistani factories, but it drives product quality and energy use in many of them. Food dryers and bakeries, pharmaceutical air handling and storage, paper machines and print halls, and the HVAC ducts that serve them all depend on knowing how much water is in the air, often at temperatures that ordinary plastic-housed room sensors were never designed for (general). A humidity transmitter that survives heat, dust and wash-down, and sends a clean 4–20 mA signal to the PLC or BMS, is what turns that air into a controlled variable.
This note looks at the Rotronic DT722 humidity and temperature transmitter, a stainless steel probe-type transmitter for relative humidity, temperature and calculated dew point in harsh environments, with probe temperatures up to 150 °C. It covers what the datasheet says about range, accuracy, outputs and construction, how to configure the order code, installation and calibration points, and when another Rotronic instrument is the better fit. Figures come from the Rotronic DT722 datasheet (59107E/2024-06) hosted on our product page and the current Tetracon product pages. General engineering advice is labelled as general, and nothing here goes beyond the manufacturer's published data. Tetracon (Pvt.) Ltd. is the sole authorized agent for Rotronic AG in Pakistan and supplies its humidity instruments with local technical support.
Why harsh-environment humidity needs a dedicated transmitter
- Heat. The DT722 probe operates from -40 to 150 °C (-40 to 302 °F), and the datasheet lists "withstands temperatures up to 150 °C" as a headline benefit (datasheet). That covers duct and process air well above normal room conditions.
- Dirt, dust and wash-down. The housing is stainless steel with IP65 (NEMA 4 level) ingress protection, and filters range from a slotted protection cap to PVDF, mesh, 5 µm and 20 µm sintered stainless steel, 2 µm metal foil and 0.7 µm oleophobic foil (datasheet).
- Stability between calibrations. Long-term stability is stated as ±1 %rh over 12 months (±1 %rh/year for the RH sensor) (datasheet), which matters when the transmitter is fixed in a duct or dryer wall and only checked during shutdowns (general).
- Listed applications. HVAC (heating, ventilation and air conditioning), the food industry, the pharmaceutical industry and the printing and paper industries (datasheet).
- The right moisture signal. Besides relative humidity, output 1 can be configured for calculated dew point, absolute humidity or wet-bulb temperature, so the control system receives the variable the process is actually controlled on (datasheet).
DT722 at a glance
| Item | Rotronic DT722 (datasheet) |
|---|---|
| Measured values | Relative humidity and temperature; dew point, absolute humidity or wet bulb calculated |
| Measurement range | 0–100 %rh; -40 to 150 °C (-40 to 302 °F) |
| Accuracy at 25 °C | < ±2 %rh (5–95 %rh); ±0.2 °C (±0.36 °F) typical |
| RH stability | ±1 %rh/year |
| RH response time | <10 s typical for 90 % of the step change |
| Calculated ranges | Dew point -40 to 60 °C; absolute humidity 0 to 200 g/m³; wet bulb -40 to 60 °C |
| Temperature output scaling | 0 to 100 °C or -40 to 150 °C; other scaling on request |
| Outputs | Output 1 (humidity value) and output 2 (temperature), 4–20 mA, max. 500 ohm |
| Supply | 8 to 32 V DC; supply voltage influence ±0.01 %rh/V typical |
| Connection | 5-pin M12; only pins 1, 2 and 4 are needed |
| Probe lengths | 200, 300, 500 or 900 mm, 19 mm (0.75 in) probe accessories |
| Operating temperature | Probe -40 to 150 °C; housing -20 to 70 °C; storage -30 to 75 °C |
| Operating humidity | 10 to 95 %rh, non-condensing (probe, housing, storage) |
| Housing and protection | Stainless steel, IP65 (NEMA 4 level) |
| Weight | 800 g (200 mm) to 1412 g (900 mm) |
| Display | With display (SD) or without (SX) |
Configuring the order code
The datasheet builds the order code from seven features in sequence (datasheet):
- A, base unit: DT722 relative humidity and temperature transmitter.
- B, display: SD with display, SX without display.
- C, moisture signal on output 1: H relative humidity; D calculated dew point (-40 to 60 °C); A calculated absolute humidity (0 to 200 g/m³); W calculated wet bulb (-40 to 60 °C).
- D, temperature scaling on output 2: T0 0 to 100 °C; TS -40 to 150 °C; TX other scaling on request.
- E, probe length: 02 (200 mm), 03 (300 mm), 05 (500 mm) or 09 (900 mm).
- F, cable: CA no cable; CB connector with screw terminal and no cable; CC connector with 2 m cable; CD connector with 5 m cable.
- G, filter: F02 slotted protection cap; F15 PVDF filter with cap; F21 mesh filter with cap; F25 or F27 sintered stainless steel 5 µm or 20 µm; F40 2 µm foil with cap; F42 0.7 µm oleophobic foil with cap.
The datasheet example DT722 + SX + A + T0 + 03 + CA + F02 is a transmitter without display, with absolute humidity and 0–100 °C temperature outputs, a 300 mm probe, no mating connector or cable, and a black slotted protection cap (datasheet). As a general rule, choose the filter for the air: an open cap for fast response in clean air, sintered or foil filters where dust or fibres are present, and the oleophobic foil where oil mist may reach the sensor (general).
Wiring and installation
- Three wires. Pin 1 (brown) is supply + 8–32 V DC, pin 2 (white) is output 2 temperature 4–20 mA and pin 4 (black) is output 1 RH 4–20 mA, each max. 500 ohm. Pins 3 and 5 are for calibration only and are not connected in service; the standard A000031 cable has four wires but only three are needed (datasheet).
- Connect both outputs. The datasheet notes that both RH and T must be connected for the temperature output to work (datasheet). Wire both loops even if the control system only uses one value.
- Mounting. A 3/4 in NPT stainless steel adjustable fitting (A000110) and an aluminium mounting flange for 19 mm stainless probes (A000100, 80 mm outside diameter) are listed accessories (datasheet). Choose the probe length so the sensor sits in representative moving air, away from walls, steam injection points and dead zones (general).
- Keep the head cooler than the probe. The probe is rated to 150 °C, but the housing is rated -20 to 70 °C (datasheet). On hot dryers and ovens, use a longer probe so the housing stays outside the hot zone and out of direct sun (general).
- Non-condensing. Operating humidity for probe, housing and storage is 10 to 95 %rh, non-condensing (datasheet). Avoid points where condensate or spray can wet the sensor, and fit a protective filter where droplets are possible (general).
Calibration and spares
- Interval. With ±1 %rh/year stated stability (datasheet), many sites check fixed humidity transmitters once a year and after any process upset, more often in GMP areas (general).
- On-site checks. Rotronic SCS humidity standards pair certified salt-solution ampoules with calibration chambers sized for 5/12/15/25 mm probes (product page); confirm a suitable chamber or adapter for the 19 mm DT722 probe before you order.
- Generators. The HygroGen2 generates 2–99 %RH and -5 to +60 °C and calibrates multiple probes at once; the HygroCal100 Advanced tests up to 7 probes of varying diameters from 5 to 95 %RH; the S904 humidity validator has 5 interchangeable probe ports and covers 10 to 90 %RH at +10 to +50 °C (product pages).
- Compatibility. DT722 transmitters are not compatible with Rotronic HC2A probes, and only the accessories and spare parts listed in the datasheet (filters, caps, cables, fittings) are compatible (datasheet). Order replacement filters by their A0000xx codes.
Notes for Pakistani plants
- Food and bakery lines. Dryers, proving rooms and cooling tunnels often run well above room temperature; the 150 °C probe rating and a dew point or absolute humidity output make the DT722 a natural fit for drying control (datasheet; general).
- Pharmaceutical HVAC. The pharmaceutical industry is a listed application (datasheet). For GMP areas, keep the transmitter in the calibration schedule with traceable certificates, as covered in our pharma cleanroom humidity guide (general).
- Paper, board and printing. Paper and printing are listed applications (datasheet). Stable hall and machine-hood humidity reduces curl, static and register problems (general).
- Summer heat and dust. Housing operating range is -20 to 70 °C and protection is IP65 (datasheet). In Sindh and Punjab summers, shade roof-top ducts and select a sintered or foil filter where dust loading is high (general).
- Not for hazardous areas. The DT722 datasheet lists no hazardous-area certification. For explosive atmospheres, use the intrinsically safe HygroFlex5-EX instead (datasheet; product page).
When another instrument is the better fit
- Rotronic HygroFlex5-EX: two-channel humidity and temperature transmitter with an intrinsically safe interchangeable stainless steel probe, ATEX / IECEx / UKEX certified, aluminium housing with optional display (product page). Choose it for Zone-classified process areas; see our HygroFlex5-EX note.
- Michell HygroSmart I7000XP: interchangeable industrial RH/temperature sensor with HygroSmart 3 technology (product page), designed for quick sensor swaps; compare it in our HygroSmart note.
- ACI DLP: configurable-span differential pressure transmitter for building and duct static pressure (product page), often installed alongside duct humidity transmitters in HVAC; see our ACI DLP note.
Related products
| Need | Product | Key point |
|---|---|---|
| Harsh-environment RH and temperature transmitter | Rotronic DT722 | 0–100 %rh, probe -40 to 150 °C, 2x 4–20 mA, stainless steel IP65 (datasheet) |
| Hazardous-area humidity | Rotronic HygroFlex5-EX | Intrinsically safe probe, ATEX / IECEx / UKEX (product page) |
| Portable probe calibrator | Rotronic HygroGen2 | 2–99 %RH, -5 to +60 °C, multiple probes (product page) |
| Multi-probe validation | Rotronic HygroCal100 Advanced | Up to 7 probes, 5–95 %RH (product page) |
| Transportable validator | Rotronic S904 | 5 probe ports, 10–90 %RH, +10 to +50 °C (product page) |
| Salt-solution humidity standards | Rotronic SCS | Certified ampoules, traceable to national standards (product page) |
For more background, read our humidity probe calibration guide and the HygroGen2 note. The wider range is on the Humidity & Pressure category page, our selection guides and the Rotronic AG manufacturer page.
What to send with an enquiry
- The measuring point: duct, dryer, oven, room, storage area or process hood, and the normal and maximum air temperature at the probe and at the housing.
- The variable you control on: relative humidity, dew point, absolute humidity or wet bulb, and the temperature scaling you need.
- Required probe length, mounting method (3/4 in NPT fitting or flange) and whether a local display is wanted.
- Dust, fibre, oil mist or wash-down conditions, so we can suggest the filter.
- The supply voltage and the input card or BMS that will read the two 4–20 mA loops.
- Your calibration plan: on-site with salt standards or a generator, or at our lab.
FAQ
What does the Rotronic DT722 measure?
Relative humidity from 0 to 100 %rh and temperature from -40 to 150 °C. The humidity output can instead be set to calculated dew point (-40 to 60 °C), absolute humidity (0 to 200 g/m³) or wet-bulb temperature (-40 to 60 °C).
How accurate is it?
At 25 °C the datasheet gives < ±2 %rh between 5 and 95 %rh and ±0.2 °C typical for temperature, with RH sensor stability of ±1 %rh per year and an RH response time under 10 seconds typical (T90).
Can the whole transmitter go into a 150 °C process?
No. The probe is rated -40 to 150 °C, but the housing is rated -20 to 70 °C. Choose a probe length (200, 300, 500 or 900 mm) that keeps the stainless steel housing outside the hot zone.
Which outputs and supply does it use?
Two 4–20 mA outputs (max. 500 ohm): output 1 for the humidity value and output 2 for temperature, powered from 8 to 32 V DC through a 5-pin M12 connector. Only pins 1, 2 and 4 are needed, and both RH and T must be connected for the temperature output to work.
Is the DT722 suitable for hazardous areas?
The DT722 datasheet lists no hazardous-area approval. For explosive atmospheres, Tetracon offers the Rotronic HygroFlex5-EX intrinsically safe humidity and temperature transmitter, which is ATEX, IECEx and UKEX certified.
Planning humidity measurement on a dryer, duct, cleanroom or warehouse in Pakistan? Contact Tetracon with the temperature, the variable you control on and the mounting details. You can also browse the DT722 product page and the Rotronic AG manufacturer page.