Humidity generator calibrator in Pakistan: Rotronic HygroGen2 humidity and temperature probe calibrator for pharma and ISO 17025 labs
Rotronic HygroGen2: humidity & temperature probe calibrator, <±0.1 %rh / <±0.01 °C, HG2-S / XL (datasheet).

Humidity and temperature probes drift. In Pakistani pharmaceutical plants, ISO/IEC 17025 calibration labs, meteorological stations and HVAC commissioning, the question is not whether to recalibrate — it is how to generate stable, traceable humidity and temperature conditions without shipping every transmitter back to a national lab. A portable humidity generator that can sit at the installation point, run multi-point sequences and accept probes from more than one manufacturer is often the practical answer.

This note looks at the Rotronic HygroGen2 humidity and temperature probe calibrator (HG2-S and HG2-XL). Figures come from the Rotronic HygroGen2-S/XL datasheet (59061E/2025-04) hosted on our product page and from the current Tetracon product pages. Where we give general engineering advice it is labelled as general. No performance claim here goes beyond the manufacturer's published data. Tetracon (Pvt.) Ltd. is the sole authorized agent for Rotronic AG in Pakistan.

Why a humidity & temperature generator/calibrator?

  • Full working-range calibration, not a single salt point. Saturated salt solutions and spot checks are useful, but many quality systems need the probe exercised across its working humidity and temperature range with documented dwell times (datasheet; general for pharma / ISO 17025 practice).
  • On-site system qualification. Being portable and robust, HygroGen2 can be set up at the installation location so the whole measurement chain — probe, cable, transmitter, DCS faceplate — is qualified together (datasheet).
  • Multi-probe throughput. Standard doors take multiple probes at once (HG2-S: up to 6 through 30 mm ports; HG2-XL: up to 19 through the door plus internal shelves), which cuts calendar time when a cleanroom or warehouse sensor fleet is due (datasheet).
  • Fast equilibrium for lab and field work. Humidity equilibrium is typically about 5 minutes on HG2-S and about 15 minutes on HG2-XL for 5…95 %rh to 0.1 %rh stability; temperature changes 23…50 °C to 0.01 °C stability follow the same order (datasheet).
  • Optional automation and chilled-mirror reference. AutoCal, Remote API and external chilled-mirror reference integration are published as licence unlocks so a lab can grow into unattended multi-point certificates and lower uncertainty when required (datasheet).

How HygroGen2 works

HygroGen2 generates a sealed chamber of mixed-flow air: a piezoelectric humidifier with digital PID control supplies moisture, an internal desiccant cell supplies dry air, and a Peltier thermoelectric element with a radial mixing fan holds temperature. At equilibrium, RH control is better than ±0.1 %rh (datasheet).

The standard temperature control range is 0…60 °C and the standard humidity range is 5…95 %rh. Optional licence extensions widen that to −5…60 °C and 2…99 %rh. A touch-screen interface, USB ports for peripherals and HygroClip2 USB cables, and a user-programmable set-point / ramp-soak function (20 programs, up to 200 set-points each) let instruments be calibrated at multiple points without constant operator intervention (datasheet).

An external heated sample loop (temperature-controlled outlet and inlet, 6 mm fittings) supports connection of a chilled-mirror reference hygrometer so the control HygroClip can be verified, or so the chamber can feed a stable humidity to an external application. Integrated Rotronic HW4 software is published as FDA 21 CFR Part 11 compliant for data acquisition and digital adjustment of compatible probes; the water reservoir includes UV sterilisation, and the desiccant cell has condition monitoring (datasheet).

Two chamber sizes are published. HG2-S: 2 litre chamber / 1.5 litre working volume; about 13 kg; 45 × 41 × 21 cm. HG2-XL: 20 litre chamber / 15 litre working volume (10× the HG2-S chamber); about 37 kg; 80 × 62 × 41 cm. Standard doors are HG2-S 6 × 30 mm ports and HG2-XL 8 × 30 mm (custom doors and probe sleeves for other manufacturers' diameters are listed). Enclosure rating is IP20; power is 110…240 V AC 50/60 Hz; operating ambient is 10–35 °C (datasheet).

HygroGen2 at a glance

ItemFrom the datasheet (59061E/2025-04)
ModelsHG2-S (2 L / 1.5 L working) and HG2-XL (20 L / 15 L working)
Stability (advantages)<±0.1 %rh; <±0.01 °C
Standard ranges0…60 °C; 5…95 %rh (optional extensions −5…60 °C; 2…99 %rh)
Humidity change to 0.1 %rh~5 min (HG2-S); ~15 min (HG2-XL) for 5…95 %rh
Temperature change to 0.01 °C~5 min (S) / ~15 min (XL) for 23…50 °C; <25 / <35 min for 23…0 °C
Thermal gradients<±0.05 °C (15…50 °C); <±0.1 °C (5…60 °C); <±0.15 °C (0…5 °C)
Typical calibration uncertainty±1.5 %rh (k=2) at 23 °C; ±0.15 °C (k=2) 15…50 °C
Control probeHygroClip2 capacitive RH + Pt100; ±1.3 %rh (0…10 and 30…50 °C); ±1.8 %rh (50…60 °C)
Programmer20 user programs; up to 200 set-points per program
Probe capacityHG2-S up to 6 through door; HG2-XL up to 19 through door + shelves
Optional featuresAutoCal; Remote API; chilled-mirror ExtRef; range extensions; remote screen share
Software / complianceIntegrated Rotronic HW4 (FDA 21 CFR Part 11 compliant) per datasheet advantages
Power / enclosure110…240 V AC 50/60 Hz; IP20; operating 10–35 °C
Weight / sizeHG2-S ~13 kg, 45×41×21 cm; HG2-XL ~37 kg, 80×62×41 cm
Rotronic HygroGen2 humidity and temperature probe calibrator with key specifications
HygroGen2: <±0.1 %rh / <±0.01 °C stability, HG2-S ~5 min equilibrium, multi-probe doors (datasheet).
Choosing between Rotronic HygroGen2, HygroCal100 Advanced and S904 humidity calibration tools
Which Rotronic humidity calibration tool for which duty, from the datasheets and our product pages.

Where it fits in Pakistan

  • Pharmaceutical quality and GxP. Probe and transmitter calibration / system qualification where FDA 21 CFR Part 11 style electronic records and multi-point sequences are expected (datasheet applications; general for local pharma QA).
  • ISO/IEC 17025 humidity calibration laboratories. The datasheet notes HygroGen2 is favoured in ISO 17025 humidity labs worldwide; the optional chilled-mirror ExtRef path supports lower uncertainty work (datasheet).
  • On-site calibration of installed humidity loops. Cleanrooms, warehouses, stability chambers and HVAC systems where removing every probe to a remote lab is costly (datasheet “on-site calibration solution”; general).
  • Meteorological and research instruments. Multi-point RH/T checks on weather and research probes using manufacturer-specific sleeves (datasheet; general).
  • Fleet calibration of Rotronic HygroClip2 / HCD probes. With AutoCal, pre-programmed set-points can calibrate and adjust HygroClip2 RH probes and generate PDF certificates (datasheet).
  • Sister tools when a full generator is not required. For a compact portable humidity generator see HygroCal100 Advanced; for a humidity validator see S904; for salt / SCS standards see SCS humidity standards. An earlier overview of the family is in Humidity probe calibration: HygroGen2, HygroCal100 and SCS.

Practical notes for a clean calibration setup

  • Pick HG2-S vs HG2-XL by throughput and probe size. HG2-S is the portable workhorse (about 13 kg, ~5 min humidity equilibrium). HG2-XL suits larger instruments, mapping, and higher probe counts (about 37 kg, ~15 min humidity equilibrium) (datasheet).
  • Order the door and sleeves for the probes you actually own. Standard 30 mm ports need B8 sleeves for smaller diameters; custom doors exist for >30 mm instruments. Clear acrylic doors are listed for HG2-S instruments with displays (datasheet).
  • Treat range extensions as options, not the default. Standard is 0…60 °C / 5…95 %rh. If you need −5 °C or 2…99 %rh, specify the TempExt / HumiExt licence codes (or a feature bundle) up front (datasheet).
  • Use the heated sample loop when uncertainty matters. Connecting a chilled-mirror reference (ExtRef licence) lets you verify the control probe or use the mirror as the AutoCal / Remote API reference; the datasheet cites sub-0.7 %rh uncertainty at ambient with a suitable mirror (datasheet).
  • Plan water and desiccant logistics. UV sterilisation addresses waterborne contaminants; desiccant condition is monitored so regeneration or replacement is scheduled, not guessed (datasheet; general for dusty Pakistani field sites).
  • Keep the unit within 10–35 °C ambient. The enclosure is IP20 — use it indoors or under cover, not as an outdoor weather station (datasheet).
  • Automate only after the method is proven. AutoCal and Remote API are powerful once set-points, dwell times and acceptance limits are written into the lab method (datasheet; general).

Related Rotronic tools and when to choose them

DutyProductFrom the datasheet / product page
Portable RH/T generator, multi-probe, pharma / ISO 17025Rotronic HygroGen2 (HG2-S / XL)<±0.1 %rh / <±0.01 °C stability; AutoCal / ExtRef options; HW4 21 CFR Part 11
Compact portable humidity generatorHygroCal100 AdvancedSmaller generator for field / bench RH checks; see family overview blog
Humidity validatorRotronic S904Validator duty when a full programmable generator is not required
SCS / salt humidity standardsSCS humidity standardsSpot checks and simple reference points
Hazardous-area humidity transmitterHygroFlex5-EXIS transmitter; see HF5-EX blog

As a rule of thumb, choose HygroGen2 when you need a programmable RH/T chamber with multi-probe throughput and optional automation / chilled-mirror reference; choose HygroCal100 Advanced or S904 when a lighter or simpler validator/generator is enough; and keep SCS standards for spot checks. For interchangeable RH sensors that swap in the field without a generator visit, see the Michell HygroSmart I7000XP note; for dryer dew-point transmitters see Michell SF52; for WiFi warehouse RH/T logging see Lascar EL-WiFi-TH+ (datasheets; general).

What to send with an enquiry

  • HG2-S vs HG2-XL preference, and how many probes you need to calibrate at once (including non-Rotronic diameters).
  • Required humidity and temperature span — standard 5…95 %rh / 0…60 °C, or optional extensions.
  • Whether AutoCal, chilled-mirror ExtRef, Remote API, screen share or a feature bundle is needed.
  • Door / sleeve requirements (30 mm standard vs custom) and whether a clear acrylic door is useful.
  • Transit bag or heavy-duty case, spare desiccant, and HygroClip2 USB calibration cables.
  • Whether sister tools (HygroCal100 Advanced, S904, SCS standards, HF5-EX) are also required on the same project.

FAQ

What is the difference between HG2-S and HG2-XL?

Chamber and working volume: HG2-S is 2 L / 1.5 L; HG2-XL is 20 L / 15 L. Humidity equilibrium to 0.1 %rh is typically about 5 minutes on HG2-S and about 15 minutes on HG2-XL. HG2-S weighs about 13 kg; HG2-XL about 37 kg. Probe capacity is up to 6 through the HG2-S door versus up to 19 through the HG2-XL door plus shelves.

What humidity and temperature ranges are standard?

Standard control ranges are 0…60 °C and 5…95 %rh. Optional licence extensions widen temperature to −5…60 °C and humidity to 2…99 %rh. Stability figures published in the advantages are better than ±0.1 %rh and better than ±0.01 °C.

Can it calibrate probes from other manufacturers?

Yes. The datasheet describes HygroGen2 as an on-site calibration solution for humidity and temperature measuring equipment generally, with probe sleeves for other manufacturers' diameters and custom doors available. AutoCal's automated adjust/certificate path is published specifically for Rotronic HygroClip2 / HCD probes.

Is HygroGen2 FDA 21 CFR Part 11 capable?

The datasheet lists integrated FDA 21 CFR Part 11 compliant Rotronic HW4 software for data acquisition and digital adjustment of compatible probes. Confirm the exact software revision (HW4 / HW5-P wording on the sheet) against the quotation for your unit.

When should I choose HygroCal100 Advanced or S904 instead?

Choose HygroCal100 Advanced for a more compact portable humidity generator, or S904 when a humidity validator is enough. Choose HygroGen2 when you need the larger programmable chamber, multi-probe doors, optional AutoCal / chilled-mirror reference and the pharma / ISO 17025 feature set described above.

Need help selecting a Rotronic humidity generator or calibrator for a Pakistani pharma plant, ISO 17025 lab or on-site HVAC qualification? Contact Tetracon with the model preference (HG2-S / XL), probe count and diameters, required RH/T span and any AutoCal / ExtRef options — or browse the HygroGen2 product page and the Rotronic AG manufacturer page.

Talk to a Tetracon engineer All articles