Home About Us Manufacturers Products Services Clients Downloads Blog Guides FAQ Contact
21 CFR Part 11 data loggers in Pakistan: Lascar EasyLog EL-21CFR-2-LCD+, EL-21CFR-TP-LCD+ and EL-WiFi-21CFR-DTC for pharma temperature and humidity records
Lascar EasyLog 21CFR range: EL-21CFR-2-LCD+ temperature and humidity logger, EL-21CFR-TP-LCD+ thermistor probe logger and EL-WiFi-21CFR-DTC dual-thermocouple WiFi sensor, for use as part of a 21 CFR Part 11 compliant system (datasheets).

A pharmaceutical store, QC laboratory or cold room in Pakistan can have accurate temperature readings and still fail an audit on the records. Readings on a USB logger that anyone can edit, export files with no record of who changed what, and printouts signed days later are the usual problems. When a quality system treats the electronic record as the official record, the logger and its software also have to control who can do what, keep an audit trail and stop data from being changed.

This note looks at the Lascar EasyLog 21CFR data loggers, a family of USB and WiFi loggers designed to be used as part of a 21 CFR Part 11 compliant system. It covers the EL-21CFR-TP-LCD and EL-21CFR-TP-LCD+ thermistor probe loggers, the EL-21CFR-2-LCD+ temperature and humidity logger and the EL-WiFi-21CFR-DTC dual-thermocouple WiFi sensor. Figures come from the Lascar datasheets hosted on our product pages and from the current Tetracon product pages. General engineering advice is labelled as general, and nothing here goes beyond the manufacturer's published data. Tetracon (Pvt.) Ltd. is the sole authorized agent for Lascar Electronics in Pakistan and supplies EasyLog data loggers with local technical support.

Why the record matters as much as the reading

  • 21 CFR Part 11 is about electronic records. The US FDA regulation sets the criteria under which electronic records and signatures are treated as equivalent to paper records (product page). Pakistani manufacturers exporting to regulated markets, and quality systems that follow the same data-integrity principles, are asked to show controls of this kind (general).
  • The logger is part of a system, not the whole answer. Lascar's datasheets say the loggers are for use as part of a 21CFR Part 11 compliant system, and the WiFi model can be installed as part of a compliant system with additional controls (datasheet). Procedures, user management and validation are still the site's job (general).
  • What the 21CFR software adds. Individual users with specific permissions, full software and session audit trails, email alerts for failed log-in attempts, digital signatures on all reports and comments on specific readings. Logged data and audit trails are stored in an encrypted format that cannot be edited (datasheet).
  • Typical duties. Finished-goods and raw-material stores, stability and sample rooms, QC laboratories, cold rooms and refrigerators, and incubators or ovens where a thermocouple is needed (product page applications; general).

EasyLog 21CFR range at a glance

ItemEL-21CFR-TP-LCD / TP-LCD+EL-21CFR-2-LCD+EL-WiFi-21CFR-DTC
MeasuresTemperature from an external thermistor probe (TP-LCD supplied with the standard EL-P-TP probe; TP-LCD+ with the high-accuracy EL-P-TP+ probe)Temperature, relative humidity and calculated dew pointTwo channels from external thermocouples; two K-type probes supplied; K, N, J and T-type compatible
Range-40 to +125 °C-35 to +80 °C; 0–100 %RH0 to +200 °C with the supplied K-type probes; K-type -270 to +1372 °C, J-type -210 to +1200 °C, T-type -270 to +400 °C with other probes
AccuracyLogger error ±0.1 °C; probe accuracy per the probe chart in the datasheet0.45 °C typical (5–60 °C); 2.05 %RH typical (10–90 %RH); dew point 1.5 °C typical (overall error)±1.5 °C display accuracy; ±2.5 °C with the supplied K-type probe
Resolution0.1 °C data, 0.5 °C display0.5 °C and 0.5 %RH0.1 °C
Memory32,510 readings16,382 temperature and 16,382 humidity readingsStreams to the 21CFR Cloud; logs locally while WiFi is lost (max 30 days at a 10-second sample interval)
Logging rate1 second to 12 hours10 seconds to 12 hours10 seconds to 12 hours; transmission every 1 minute to 24 hours
Display and alarmsHigh-contrast LCD (current, max, min); red/green LEDs; user-set high and low alarms with hold and delayed alarm optionsLCD with current, max and min for both channels; one LED per channel; high and low alarms for temperature and humidityDisplay with configurable high and low alarms and indicator; email alerts via the 21CFR Cloud
Power3.6 V 1/2AA lithium battery, about 2 years at 25 °C, 1-minute logging, LCD on3.6 V 1/2AA lithium battery, about 2 years at 25 °C, 1-minute logging, LCD onRechargeable battery, more than 6 months; can be powered permanently by a USB adapter
Operating range and protectionLogger -35 to +80 °C-35 to +80 °C; IP67 with the cap fittedLogger -20 to +60 °C; IP40, probe tip IP67
SoftwareEasyLog 21CFR PC software, free download (Windows 7, 8 and 10, 32 and 64-bit)EasyLog 21CFR WiFi Sensor Software (free) plus a 21CFR Cloud subscription
Size135 x 24 x 21 mm126 x 25 x 22 mm82 x 70 x 36 mm (excluding bracket)
Lascar EL-21CFR-2-LCD+ temperature and humidity data logger with key datasheet figures for 21 CFR Part 11 records
Lascar EL-21CFR-2-LCD+: temperature, humidity and dew point logging with EasyLog 21CFR software controls (datasheet figures).
Lascar EasyLog 21CFR model selection table: EL-21CFR-TP-LCD, EL-21CFR-TP-LCD+, EL-21CFR-2-LCD+ and EL-WiFi-21CFR-DTC with ranges, memory and logging rates
Choosing between the four EasyLog 21CFR models by measurement, range, memory and logging rate (datasheets).

Choosing the model

  • Refrigerators, cold rooms and freezers: EL-21CFR-TP-LCD or TP-LCD+. The probe goes inside the cold space while the logger with its display stays outside, and the -40 to +125 °C range covers fridge, freezer and warm-room duties (datasheet; general). Choose the TP-LCD+ when you want the high-accuracy EL-P-TP+ “Type A” thermistor probe supplied; it is sold separately for the TP-LCD (datasheet).
  • Stores, warehouses, stability and sample rooms: EL-21CFR-2-LCD+. Where the specification is temperature and relative humidity, such as a 15–25 °C store with a humidity limit, one logger records both and the software calculates dew point (datasheet; general).
  • Ovens, incubators, autoclave rooms and process checks: EL-WiFi-21CFR-DTC. Two thermocouple channels, interchangeable probe types up to +1372 °C with K-type, and data streamed over WiFi instead of collected by hand (datasheet; general).
  • Several rooms on one site. USB loggers need someone to collect and download them; the WiFi sensor sends data to the 21CFR Cloud with alerts, which suits sites with reliable 2.4 GHz WiFi (datasheet; general).
  • No Part 11 requirement? The standard EasyLog equivalents, such as the EL-USB-2-LCD+ and EL-WiFi-TH+, use the regular EasyLog software and cloud without the 21CFR controls (datasheet; general).

EasyLog 21CFR software and 21CFR Cloud

  • Set-up. Logger name, °C or °F, logging rate, high and low alarms, immediate or delayed start, and probe type on the TP models; push-to-start logging is also available (datasheet).
  • Data control. User permissions, full software and session audit trails, failed log-in email alerts, digital signatures on all reports and comments on individual readings; data cannot be edited (datasheet).
  • Reports. Graph and annotate data, or export to Excel, PDF or JPEG (datasheet).
  • 21CFR Cloud for the WiFi model. Permission-based use and access, authority-based sign-off on actions, records that cannot be edited or deleted and a complete system audit trail, with event logs and email alerts. The EL-WiFi-21CFR-DTC works only with the 21CFR Cloud, not the standard EasyLog Cloud, and needs a subscription (datasheet).
  • Validation is still needed. Plan installation and operational checks, user roles and a review procedure for the software as part of your computerised-system validation (general).

Practical set-up tips for Pakistan

  • Map before you mount. Place loggers where a temperature-mapping study found the warmest and coldest points, not just next to the door (general).
  • Plan for power cuts. The USB loggers run on their own battery and keep logging through load-shedding. For the WiFi sensor, the battery lasts more than 6 months, but if the router goes down it logs locally for up to 30 days at a 10-second sample interval, so keep the router on a UPS (datasheet; general).
  • Replace batteries on a schedule. Lascar recommends replacing the battery annually or before logging critical data. Stored readings are kept when the battery is changed, but logging stops until the logger is restarted from the 21CFR software (datasheet).
  • Do not leave a USB logger plugged in. Leaving it in the PC's USB port for long periods drains battery capacity (datasheet).
  • Protect the humidity sensor. Keep the EL-21CFR-2-LCD+ away from solvents and strong-smelling chemicals; the humidity sensor needs about 20 minutes to reach 90 % of a change, and up to four hours to settle after a step change during calibration (datasheet).
  • Calibration certificates. Lascar offers a traceable calibration certificate service for its temperature data loggers, and we can advise on in-house checks with a Rotronic HygroGen2 humidity generator for RH loggers (datasheet; general).

Where EasyLog 21CFR loggers fit in Pakistan

  • Pharmaceutical manufacturers for raw-material, finished-goods and stability stores, and for export-oriented plants whose customers audit electronic records (general).
  • QC and microbiology laboratories for reagent fridges, incubators and sample storage (product page applications; general).
  • Distributors, hospitals and blood banks with cold rooms and medical refrigerators that need traceable records (product page applications; general).
  • Food and nutraceutical plants with GMP-style quality systems (product page applications; general).

Related products and when to choose them

DutyProductKey figure (datasheet / product page)
Fridge, freezer or cold-room probe logging with Part 11 controlsLascar EL-21CFR-TP-LCD-40 to +125 °C, 32,510 readings, 1 s to 12 h logging (datasheet)
Same duty with the high-accuracy probe suppliedLascar EL-21CFR-TP-LCD+EL-P-TP+ Type A thermistor probe supplied (datasheet)
Store and lab temperature and humidity with Part 11 controlsLascar EL-21CFR-2-LCD+-35 to +80 °C, 0–100 %RH, 0.45 °C and 2.05 %RH typical, IP67 with cap (datasheet)
Ovens, incubators and process temperatures over WiFiLascar EL-WiFi-21CFR-DTCDual thermocouple, 0 to +200 °C with supplied K probes, 21CFR Cloud (datasheet)
Temperature and humidity logging without Part 11 controlsLascar EL-USB-2-LCD+-35 to +80 °C, 0–100 %RH, standard EasyLog software (datasheet)
WiFi room monitoring on the standard EasyLog CloudLascar EL-WiFi-TH+-20 to +60 °C, 0–100 % RH, IP55 (datasheet)
Vaccine fridges with glycol-bottle probeLascar EL-WiFi-VACX-40 to +60 °C, calibration certificate at 2 and 8 °C (datasheet)

For more background, read our vaccine and pharma cold-chain monitoring guide, the pharma cleanroom humidity and temperature guide, the EL-WiFi-TH+ note and the humidity probe calibration note. The wider range is on the Environmental, Humidity & Pressure and Process Instruments category pages, our selection guides and the Lascar Electronics manufacturer page.

What to send with an enquiry

  • The locations to monitor (fridge, freezer, cold room, store, lab, incubator or oven) and the number of points.
  • The parameters and limits: temperature only, or temperature and humidity, and the alarm limits.
  • Whether USB download is acceptable or data must stream over WiFi, and whether site WiFi is 2.4 GHz with enterprise security.
  • Your quality-system requirement (21 CFR Part 11 or similar data-integrity controls) and who will validate the system.
  • Whether traceable calibration certificates are required with delivery.

FAQ

Are the Lascar EasyLog 21CFR loggers 21 CFR Part 11 compliant?

They are designed to be used as part of a 21 CFR Part 11 compliant system. The EasyLog 21CFR software provides user permissions, audit trails, digital signatures and encrypted data that cannot be edited, and the 21CFR Cloud adds permission-based use, authority sign-off and a full system audit for the WiFi model. Compliance also depends on your procedures and validation.

Which model should I use for a pharmaceutical store?

For temperature and humidity in a store or warehouse, the EL-21CFR-2-LCD+ measures -35 to +80 °C and 0–100 %RH, with 0.45 °C and 2.05 %RH typical accuracy and calculated dew point. For fridges and freezers, use the EL-21CFR-TP-LCD or TP-LCD+ with the probe inside the cold space.

How many readings do they store?

The EL-21CFR-TP-LCD and TP-LCD+ store 32,510 readings, and the EL-21CFR-2-LCD+ stores 16,382 temperature and 16,382 humidity readings. The EL-WiFi-21CFR-DTC streams to the 21CFR Cloud and logs locally for up to 30 days at a 10-second interval if WiFi is lost.

What happens during a power cut?

The USB loggers run on a 3.6 V lithium battery that typically lasts 2 years, so they keep logging. The WiFi sensor has a rechargeable battery rated at more than 6 months and keeps readings while the network is down, then uploads them when it reconnects.

Is the software free?

Yes. EasyLog 21CFR PC software and EasyLog 21CFR WiFi Sensor Software are free downloads for Windows. The WiFi model also needs a 21CFR Cloud subscription.

Setting up 21 CFR Part 11 temperature or humidity records in Pakistan? Contact Tetracon with the locations, limits and quality-system requirements. You can also browse the EL-21CFR-2-LCD+ product page and the Lascar Electronics manufacturer page.

Talk to a Tetracon engineer All articles