Every nitrogen generator, PSA oxygen plant and inerted vessel in a plant has an online oxygen analyser on it, and the whole system trusts that one reading. When the number looks wrong, or when a customer, auditor or QA team asks for proof that the gas is really what the panel says, you need a second, independent measurement that you can carry to the point of use and connect in minutes. That is the job of a transportable verification analyser.
This note looks at the Ntron Yellow Box portable oxygen analyzer, a carry-case zirconia instrument offered as models 100, 200 and 300 for ppm and percent oxygen. Figures come from the Yellow Box datasheet (NT-DS-2104-EN-01, 06/2024) hosted on our product page and from the current Tetracon product pages. Where we give general engineering advice it is labelled as general. Nothing here goes beyond the manufacturer's published data. Tetracon (Pvt.) Ltd. is the sole authorized agent for Ntron Gas Measurement in Pakistan.
Why carry a second oxygen analyzer at all?
- Verification of primary instrumentation. Ntron lists this as the first Yellow Box application: a line-powered reference you take to an installed analyser to confirm its reading (datasheet).
- N2 and O2 generators. Commissioning, acceptance and periodic checks of on-site generators, where the purity figure is the contract figure (datasheet; general).
- Industrial, medical and pharmaceutical gas. The same three application areas appear on the datasheet, where a documented independent check supports quality records (datasheet; general).
- One case for ppm and percent. The datasheet describes three discrete, independent sensors so a single instrument covers trace ppm and percent oxygen, with ranges from 0–1000 ppmV up to 0–96 % O2 depending on model (datasheet).
- Built for transport. A high-impact co-polymer polypropylene case that is IP67 when closed for transportation (datasheet).
How the Yellow Box measures
The analyser uses Ntron's long-life, quick-response zirconia technology, the same engineering behind the Ntron Microx (datasheet). Sample gas enters through a bulkhead compression fitting under positive pressure, passes the internal regulation (up to 2 barg / 29 psig internal preset), and the sample pressure and flow are set at 0.03 MPa (0.3 barg / 5 psig) and 0.55 SCFH (250 ml/min) before it reaches the sensors; a flow meter on the sample-out port shows that gas is actually flowing (datasheet).
Each analyser channel has its own 4–20 mA output, and the front panel carries an analyzer select switch and 4–20 mA test points for each channel (datasheet). In practice the test points are useful when you want to compare the portable reading against the plant DCS or a recorder rather than just reading the display (general).
Ntron Yellow Box at a glance
| Item | From the datasheet |
|---|---|
| Sensor | Zirconia (ZR) |
| Measurement range, model 100 | 0…1000 ppmV, 0…25 % |
| Measurement range, model 200 | 0…1000 ppmV, 0…25 % |
| Measurement range, model 300 | 0…1000 ppmV, 0…25 %, 0…96 % |
| Output resolution / lower detection limit | 1 ppmV or 0.01 % |
| Response time (T90) | < 15 seconds at 25 °C (77 °F) within selected range |
| Pressure range | Up to 2 barg (29 psig) internal preset |
| Operating temperature | −25…+60 °C (−13…+140 °F) |
| Humidity | 0…95 %rh non-condensing |
| Sensor life expectancy / shelf life | Up to 5 years / unlimited |
| Output signal (per analyser channel) | 4…20 mA |
| Power supply | 85…264 V AC (50 / 60 Hz); battery-powered option on model 100 |
| Maximum power consumption | 8 W (100), 4.8 W (200), 9.6 W (300) |
| Sample connections | Sample in: 6 mm or 1/4" bulkhead tube compression fittings |
| Ingress protection / case | IP67 when case closed for transportation; high-impact co-polymer polypropylene |
| Dimensions a × b × c | 410 × 300 × 165.5 mm (100); 350 × 300 × 150 mm (200 and 300) |
| Compliance | EMC: EN 50270 |
Choosing between model 100, 200 and 300
- Model 300 is the only version with the 0…96 % range in addition to 0…1000 ppmV and 0…25 %, so it is the one to choose when you must verify oxygen concentrators or high-purity oxygen as well as nitrogen (datasheet; general).
- Models 100 and 200 cover 0…1000 ppmV and 0…25 %, which spans trace oxygen in nitrogen and inert gas through to air-level oxygen (datasheet; general).
- Model 100 is marked as the battery-powered option and is the larger case (410 mm wide against 350 mm), which matters if checks are made away from a mains socket, for example on a remote skid or a gas cylinder store (datasheet; general).
Practical notes for a trustworthy verification
- Tap the same point as the online analyser. Connect the Yellow Box to the same sample line, or a tee next to it, so both instruments see the same gas (general).
- Respect the pressure limits. The analyser regulates internally up to 2 barg (29 psig); higher-pressure lines need an external regulator before the sample-in port (datasheet; general).
- Confirm flow before reading. Check the sample-out flow meter, then allow the reading to settle; the datasheet T90 is under 15 seconds at 25 °C within the selected range (datasheet; general).
- Purge the sample tubing for ppm work. At ppm oxygen levels, air trapped in new tubing or a loose fitting will read high; use clean metal tubing with compression fittings and purge before you record a figure (general).
- Pick the right range. Use the ppm channel for nitrogen and inert gas purity and the percent channels for air, enriched oxygen or concentrator outlets (datasheet; general).
- Record both readings. Note the Yellow Box reading, the online analyser reading, time and sample point; the 4–20 mA test points let you log the portable value with a separate recorder if needed (datasheet; general).
Where it fits in Pakistan
- On-site nitrogen generators in food packaging, pharmaceutical, electronics and laser-cutting plants, where the PSA or membrane purity figure needs independent checks (general).
- Hospital PSA oxygen plants, where a model 300 can verify the 0–96 % range against the plant's own monitor (datasheet range; general application).
- Industrial gas fillers and distributors checking cylinder and bulk gas quality before dispatch (datasheet application; general).
- Service teams and instrumentation contractors who maintain online oxygen analysers at several sites and want one transportable reference (general).
Related products and when to choose them
| Duty | Product | Key figure (datasheet / product page) |
|---|---|---|
| Transportable verification of O2 analysers and generators | Ntron Yellow Box | 0–1000 ppmV to 0–96 % O2 by model; T90 < 15 s; IP67 case closed (datasheet) |
| Fixed oxygen analyser with remote sensor for generators and glove boxes | Ntron Microx | ZR 0–1,000 ppmV, 0–25, 0–96 %; 3 relays; 4–20 mA and RS232 (product page) |
| Wall-mounted O2 analyser with alarm beacon for inert gas lines | Ntron Microx-OL | 0–1,000 ppm, 0–25 % or 0–96 % O2; internal pressure and flow regulation (product page) |
| Medical gas quality: O2, CO2, CO and optional moisture | Ntron MoGas | O2 zirconia 0–96 %; CO2 0–1000 ppm; CO 0–50 ppm (product page) |
| Portable spot checks of trace ppm oxygen | AII GPR-1100 / GPR-1200 | 0–10 ppm up to 0–10,000 ppm O2, four ranges (product page) |
Choose the Yellow Box when the task is checking other instruments or generators at several points with one transportable case. Choose a Microx or Microx-OL when the measurement should stay installed on the generator outlet, and MoGas when a medical-gas line needs several parameters at once — see our MoGas PSA plant note. For portable trace-ppm spot checks down to a 0–10 ppm range, the AII GPR-1100 / GPR-1200 is covered in our portable trace oxygen note. The wider range is on the Oxygen Analysis category page and the Ntron and Analytical Industries manufacturer pages.
What to send with an enquiry
- What you want to verify: nitrogen generator, oxygen generator or concentrator, installed analyser, cylinder or bulk gas.
- Expected oxygen level and the range you need (ppm, up to 25 %, or up to 96 %), which decides between model 100/200 and model 300.
- Sample pressure and temperature at the connection point, and the background gas.
- Whether checks are made near mains power or need the battery-powered option (model 100).
- Tube size at the sample point (6 mm or 1/4") and whether you need the 4–20 mA output for logging.
- Quantity, and any fixed analysers to be supplied or checked at the same time.
FAQ
What is the Ntron Yellow Box used for?
It is a compact, transportable zirconia oxygen analyzer for high-accuracy oxygen measurement and for verifying primary instrumentation. The datasheet lists N2 and O2 generators, industrial gas, medical gas and pharmaceutical applications.
What ranges does the Yellow Box cover?
Models 100 and 200 measure 0…1000 ppmV and 0…25 % O2. Model 300 adds 0…96 % O2. Output resolution and lower detection limit are 1 ppmV or 0.01 %.
How fast does it respond?
The datasheet gives a T90 response time of less than 15 seconds at 25 °C within the selected range.
Can it run without mains power?
The standard supply is 85…264 V AC (50/60 Hz). The datasheet marks model 100 as the battery-powered option, so choose model 100 if checks are made away from a mains socket.
What sample pressure does it need?
The sample must be under positive pressure. Internal regulation is preset for up to 2 barg (29 psig), and the sample is set at 0.3 barg (5 psig) and 250 ml/min. Use an external regulator on higher-pressure lines.