Condumax II Hydrocarbon Dew-Point Analyzer
Fully automatic, chilled-mirror hydrocarbon and water dew-point analyzer for natural gas quality.

Globally certified explosion/flameproof transmitter for gas and liquid moisture in hazardous areas.
The Easidew PRO XP is a robust explosion/flameproof dew-point transmitter for trace moisture measurement in gases or non-polar liquids in hazardous areas. Unlike the intrinsically safe Easidew PRO I.S., it uses Ex d flameproof/explosion-proof protection, giving OEMs and refineries a single rugged unit with global explosion-proof certification for any installation location.
Michell’s ceramic metal-oxide moisture sensor sits inside a flameproof (Ex d) enclosure rated to contain an internal explosion, with dual conduit entry for electrical connections. A secondary RS485 communications system allows on-site re-ranging and re-scaling between dew point, ppmV and ppmW outputs; an optional integral 4-digit LED display shows the configured output locally.
| Measurement Range | −110 to +20°C dp (−100 to +20°C dp variant); liquids: 0–1000 ppmW |
|---|---|
| Accuracy | ±1°C dp (+20 to −60°C); ±2°C dp (−60 to −110°C) |
| Response Time | 5 min to T95 (dry to wet) |
| Repeatability | 0.5°C dp |
| Calibration | Traceable 13-point certificate |
| Output | 4-20 mA, 2-wire, user configurable over range |
|---|---|
| Scaled Range (Gas) | Dew point: −110 to +20°C; Moisture: 0–3000 ppmV |
| Scaled Range (Liquid) | 0–1000 ppmW, factory configured per application |
| Supply Voltage | 14–28 Vdc |
| Load Resistance | Max 250Ω @14V (500Ω @24V) |
| Current Consumption | 23 mA max |
| Operating Temp | −40°C to +60°C |
|---|---|
| Compensated Range | −20°C to +50°C |
| Operating Pressure | 45 MPa (450 barg) max; 13.2 MPa (132 barg) if CRN-registered for Canada |
| Flow Rate | 1–5 Nl/min (sample block); 0–10 m/s direct insertion (gas); 0.1–0.3 L/min or 0.1–1 m/s (liquid) |
| Housing | Aluminium, epoxy/polyurethane powder-coated (standard); 316 SS optional |
|---|---|
| Ingress Protection | IP66 (NEMA 4) |
| Process Connection | 3/4" – 16 UNF, recessed Viton O-ring, 316 SS |
| Electrical Connection | Dual 3/4" NPT gland |
| Weight | 1.6 kg (aluminium); 2.4 kg (316 SS) |
| ATEX / UKEX | II 2 G D Ex db ia IIC T6 Gb, Ex tb IIIC T80°C Db, IP66 |
|---|---|
| IECEx | Ex db ia IIC T6 Gb, Ex tb IIIC T80°C Db |
| cQPSus (USA) | Class I Div.1 Groups A-D T6; Class I Zone 1 AEx db ia IIC T6 Gb |
| CRN (Canada) | All provinces, max 132 barg |
All Tetracon-supplied instruments carry the original manufacturer warranty (typically 12–24 months) plus first-line local support from our Islamabad service centre.
Yes. Our manufacturer-trained engineers offer on-site installation, commissioning and startup training anywhere in Pakistan.
Common spares are held in local stock. Calibration is available at our lab or on-site with NIST-traceable reference standards.
Stocked items ship within 3–5 working days; engineered configurations typically 6–10 weeks depending on manufacturer schedule.
Complete the form below and one of our engineers will respond within 1 working day with pricing and lead time. Prefer email? Contact our sales team.