Home About Us Manufacturers Products Services Clients Downloads Blog Guides FAQ Contact
NT Ntron Gas Measurement Ltd.

AM Trace Portable O2 & Moisture Analyzer

Portable dual O2 and moisture (dew point) analyser for additive-manufacturing build chambers.

ModelAM Trace
SKUNTR-AMTRACE
CategoryOxygen & Combustion Analyzers

Key Features

  • Portable dual O2 & H2O (dew point) analyser for AM build chambers
  • Oxygen: dual range 0–1,000 ppm / 0–25 %
  • Moisture: -60 °C to +20 °C dew point (SF82 sensor)
  • 7" colour touchscreen with onboard logging and trending
  • Internal sample pump and electronic flow sensor
  • 2 x 4–20 mA outputs, RS-485, USB / flash drive
  • Unaffected by AM off-gases including hydrogen

Overview

AM Trace is a transportable dual-measurement (O2 and H2O) analyser designed for additive manufacturing. Its electrochemical oxygen sensor measures from 0.1 ppm to 25 % O2 and is unaffected by build-chamber off-gases including hydrogen, while the SF82 moisture sensor responds quickly from -60 °C dew point.

Working Principle

An internal pump extracts sample from the build chamber past an electrochemical oxygen sensor (dual range 0–1,000 ppm / 0–25 %) and a Michell SF82 dew-point sensor (-60 to +20 °C dp). Results are shown, trended and logged on a 7" colour touchscreen and output as 2 x 4–20 mA and RS-485.

Benefits

  • One portable unit checks both O2 and moisture in the build chamber
  • Status notifications on the touchscreen
  • Portable case with handle

Technical Specifications

Oxygen
SensorElectrochemical
RangeDual range 0–1,000 ppm / 0–25 %
Accuracy±2 % of reading
Repeatability±0.5 % of reading
Response Time (T90)< 15 s
Moisture
Range-60 °C to +20 °C dew point
Accuracy±2 °C dew point
Response (T95)< 1 min (-60 to -20 °C dp)
Electrical
Power Supply24 VDC ±10 %
Outputs2 x 4–20 mA (O2 / moisture), 2 relays per channel, RS-485
Physical
Process Connection1/8" BSP input / exhaust
Dimensions350 x 253.4 x 314 mm (excl. handle)
Weight8.5 kg
Ambient Temperature5 °C to 40 °C

Downloads

Datasheet (PDF)PDF

Frequently Asked Questions

What warranty is offered on this product?

All Tetracon-supplied instruments carry the original manufacturer warranty (typically 12–24 months) plus first-line local support from our Islamabad service centre.

Do you provide installation and commissioning?

Yes. Our manufacturer-trained engineers offer on-site installation, commissioning and startup training anywhere in Pakistan.

Are spares and calibration available locally?

Common spares are held in local stock. Calibration is available at our lab or on-site with NIST-traceable reference standards.

How long is the typical lead time?

Stocked items ship within 3–5 working days; engineered configurations typically 6–10 weeks depending on manufacturer schedule.

Request a Quote — AM Trace Portable O2 & Moisture Analyzer

Complete the form below and one of our engineers will respond within 1 working day with pricing and lead time. Prefer email? Contact our sales team.