Zirconia Oxygen Analyzer System
Zirconia (ZrO2) oxygen measurement system: 0.1–25% or 0.1–100% O2, 0.5% O2 accuracy after calibration.

High-accuracy oxygen analyzers for industrial gases, from low-ppm trace O2 (GPR-1600) through percent (GPR-2600) to purity (GPR-3100).
High-accuracy oxygen analyzers for industrial process gases from low-ppm trace O2 (GPR-1600) through percent (GPR-2600) to oxygen purity (GPR-3100). Common HMI across the range; bench, rack, panel or wall mount.
All three analyzers use AII galvanic oxygen sensors, which produce zero output in the absence of oxygen and are linear up to 100 %, so only a span calibration is normally required. The GPR-1600 has an integral bypass valve to protect the trace sensor; the GPR-3100 adds a temperature-controlled sample system for an isothermal sample. Barometric pressure and temperature compensation are standard.
| GPR-1600 | 0–10, 0–100, 0–1000 ppm, 0–1 %; LDL 50 ppb; T90 < 10 s |
|---|---|
| GPR-2600 | 0–1, 0–5, 0–10, 0–25 %; LDL 50 ppm; T90 < 13 s |
| GPR-3100 | 0–100 % plus suppressed-zero 50/80/90–100 %; accuracy ±0.1 % after calibration with 95–100 % O2 |
| Accuracy | < ±2 % of selected range at constant conditions |
| Linearity | < 0.5 % of scale |
| Inlet Pressure | 0.34–2 barg (5–30 psig) with atmospheric vent |
|---|---|
| Flow Rate | 0.5–1.0 Nl/min (1–2 SCFH) |
| Gas Connections | 1/4" compression tube fittings |
| Display | Graphical LCD 12.7 × 7 cm |
|---|---|
| Analog Output | 4–20 mA isolated, 0–1 V, 0–5 V |
| Range ID | 1–5 V or 4–20 mA, optional relay contacts |
| Communications | RS485, RS232 or USB |
| Alarms | Two user-adjustable alarm relays |
| Power | Universal 100–240 V AC |
| Operating Temperature | GPR sensor 5–45 °C; XLT sensor -10–45 °C |
|---|---|
| Enclosure | Painted aluminum; wall, bench, panel or 19" rack |
All Tetracon-supplied instruments carry the original manufacturer warranty (typically 12–24 months) plus first-line local support from our Islamabad service centre.
Yes. Our manufacturer-trained engineers offer on-site installation, commissioning and startup training anywhere in Pakistan.
Common spares are held in local stock. Calibration is available at our lab or on-site with NIST-traceable reference standards.
Stocked items ship within 3–5 working days; engineered configurations typically 6–10 weeks depending on manufacturer schedule.
Complete the form below and one of our engineers will respond within 1 working day with pricing and lead time. Prefer email? Contact our sales team.