Home About Us Manufacturers Products Services Clients Downloads Blog Guides FAQ Contact
LE Lascar Electronics Ltd.

EL-USB-4 4-20mA Current Loop Data Logger

Standalone USB logger for recording 4-20mA d.c. current loop signals.

ModelEL-USB-4
SKULSC-ELUSB4
CategoryProcess Instrumentation

Key Features

  • Logs 4-20mA d.c. current loop signals directly via screw terminals
  • Stores over 32,000 readings
  • User-selectable logging rate: 1 second to 12 hours
  • Immediate or delayed logging start
  • Programmable alarm thresholds
  • Free EasyLog USB software for graphing, printing and export
  • Supplied with 1/2 AA battery and wall mount clip

Overview

The EL-USB-4 is a standalone data logger that stores more than 32,000 readings over a 4-20mA d.c. measurement range, making it well suited to trend-recording process current loop signals from transmitters and controllers. The logger is supplied with a cap featuring two screw terminals and measurement leads terminating in crocodile clips for quick connection to the monitored circuit. Data is retrieved by plugging the unit into a PC USB port and using the free EasyLog software, where readings can be graphed, printed and exported for analysis.

Working Principle

Two screw-terminal inputs connect directly into a 4-20mA current loop. The logger samples the loop current at a user-selectable interval and stores each reading with a timestamp to onboard memory, supporting both immediate and delayed logging starts so recording can be scheduled to begin at a specific time.

Benefits

  • Direct current-loop connection avoids need for separate signal conditioning
  • Delayed-start logging supports scheduled process trials
  • Long 3-year battery life at typical logging rates suits extended monitoring campaigns
  • Compact form factor for easy field or panel mounting

Technical Specifications

Measurement
Range4 to 20 mA d.c.
Internal Resolution0.05 mA d.c.
Accuracy±1%
Input Impedance110 Ω
Readings Stored32,510
Logging
Logging Rate1 second to 12 hours, user selectable
Start ModesImmediate or delayed logging start
Power & Environmental
Battery Life3 years (at 25°C, 2 second logging rate)
Battery Type1/2 AA (supplied)
Operating Temperature-35 to +80°C (-31 to +176°F)
Physical
Dimensions98 × 26.8 × 26.8 mm (3.85 × 1.05 × 1.05")

Downloads

Datasheet (PDF)PDF

Frequently Asked Questions

What warranty is offered on this product?

All Tetracon-supplied instruments carry the original manufacturer warranty (typically 12–24 months) plus first-line local support from our Islamabad service centre.

Do you provide installation and commissioning?

Yes. Our manufacturer-trained engineers offer on-site installation, commissioning and startup training anywhere in Pakistan.

Are spares and calibration available locally?

Common spares are held in local stock. Calibration is available at our lab or on-site with NIST-traceable reference standards.

How long is the typical lead time?

Stocked items ship within 3–5 working days; engineered configurations typically 6–10 weeks depending on manufacturer schedule.

Request a Quote — EL-USB-4 4-20mA Current Loop Data Logger

Complete the form below and one of our engineers will respond within 1 working day with pricing and lead time. Prefer email? Contact our sales team.