Home About Us Manufacturers Products Services Clients Downloads Blog Guides FAQ Contact
EMS EMS Kontrol

BT-3X1 / 3X2 Differential Pressure Transmitter

Low-range differential pressure transmitter, ±125 Pa to ±500 Pa, for filters, ducts and cleanrooms.

ModelBT-3X1 / 3X2
SKUEMS-BT-3X1-3X2
CategoryHumidity & Pressure

Key Features

  • Accurate and precise measurement
  • Analog (0–10 V / 4–20 mA) or Modbus RTU output
  • Long operating life
  • Easy assembly
  • Bidirectional ranges available
  • IP67 plastic enclosure (sensor excluded)

Overview

The BT-3X1 / 3X2 measures the pressure difference between two points — across a filter, between a cleanroom and its corridor, or between a duct and the room. The BT-3X1 covers 0–50 Pa, 0–100 Pa or −125…+125 Pa and the BT-3X2 covers 0–250 Pa, 0–500 Pa or −500…+500 Pa. Output is 0–10 V (BT-301/302), 4–20 mA (BT-341/342) or Modbus RTU (BT-361/362).

Working Principle

A differential pressure sensor connected to two pressure ports measures the difference between them; the transmitter converts it into the selected analog signal or a Modbus RTU value on RS485.

Benefits

  • Shows when filters need changing, before airflow drops
  • Low pascal ranges suit cleanroom and isolation-room pressure cascades
  • Standard outputs for BMS alarms and trending

Technical Specifications

Measurement
Range — BT-3X10–50 Pa, 0–100 Pa or −125…+125 Pa
Range — BT-3X20–250 Pa, 0–500 Pa or −500…+500 Pa
Resolution± 1 Pa
Accuracy± 3 %
Electrical & outputs
Supply voltage12–24 V DC
Voltage outputBT-301 / 302: 0–10 V / 2–10 V / 0–3 V
Current outputBT-341 / 342: 4–20 mA / 0–20 mA
Digital outputBT-361 / 362: Modbus RTU (RS485)
Physical
EnclosureIP67 plastic case (sensor excluded)
Dimensions (A × B × C)96 × 58 × 35 mm
AccessoriesDifferential pressure probe set and filters available from EMS Kontrol

Downloads

Datasheet (PDF)PDF
User manual (PDF)PDF

Frequently Asked Questions

What warranty is offered on this product?

All Tetracon-supplied instruments carry the original manufacturer warranty (typically 12–24 months) plus first-line local support from our Islamabad service centre.

Do you provide installation and commissioning?

Yes. Our manufacturer-trained engineers offer on-site installation, commissioning and startup training anywhere in Pakistan.

Are spares and calibration available locally?

Common spares are held in local stock. Calibration is available at our lab or on-site with NIST-traceable reference standards.

How long is the typical lead time?

Stocked items ship within 3–5 working days; engineered configurations typically 6–10 weeks depending on manufacturer schedule.

Request a Quote — BT-3X1 / 3X2 Differential Pressure Transmitter

Complete the form below and one of our engineers will respond within 1 working day with pricing and lead time. Prefer email? Contact our sales team.